didoEASY MR / didoEASY+ MR / didoEASY++ MR
didoEASY MR / didoEASY+ MR / didoEASY++ MR
1X-Ray Multi-Meter Series for Service and QA in R&F, Dental and Mammography
QUART didoEASY MR | Basic Configuration |
QUART didoEASY+MR | Added kV Measurement |
QUART didoEASY++MR | Added kV and Direct-HVL Measurement |
The QUART didoEASY meters are designed for users who emphasise high precision in dosimetric applications but do not require the performance of a full-range dosemeter package.
In basic configuration QUART didoEASY meters can be used to measure parameters which are essential for service and quality assurance operations at x-ray equipment such as dose, dose rate and time. Of course, as with all QUART meters – with maximum precision.
- kVp and PPV measurement are available in the QUART didoEASY+ MR version.
- Direct-HVL and kVp / PPV measurement are additionally available in the QUART didoEASY++ MR version.
Since the meters do not require any pre-setting procedure, measurement results are acquired very quickly:
1. Position the detector and switch on the didoEASY meter...
2. Set the x-ray equipment to the desired parameters...
3. Expose... - and simply read all measurement information from the meter‘s display.
The R detector automatically compensates all major medical x-ray beam qualities. No further corrections or compensations are required.
The M detector for mammography automatically compensates for Mo/Mo beam quality. Corrections factors are provided for further radiation qualities.
Like the QUART dido2000 series meters, the detector of the QUART didoEASY meter series also measures the integrated dose-width product (DWP) at dental panoramic equipment.* A feature that provides real added value for QUART users.
Technical specifications
General Features
- Exposures: one to acquire all parameters
- Mode: auto-start, auto-stop & auto-reset
- Power supply: Rechargeable Li Ion accu
- Battery time: approx. 120 hrs. until recharge required
- Base size: 17x7x4.5cm (LxWxH)
- System weight: 200 g
- Detector size: 6.0x1.8x0.5cm (LxWxH)
- Detector weight: negligible (!)
Specifications
didoEASY MR basic configuration
- Dose: 0.2 µGy - 999 Gy (Uncertainty: ±5% @ 25-160 kV / Resolution: 0.01 nGy)
- DLP*: 0.2 µGy*cm - 999 Gy*cm (Uncertainty: ±5% @ 50-160 kV / Resolution: 0.01 nGy*cm)
- Dose Rate: 0.3 µGy/s - 1 Gy/s (Uncertainty: ±5% over full range / Resolution: 0.1 nGy/s)
- Time: 0.5 ms - 300 s (Uncertainty: ± 0,5% / Resolution: 0.1 ms)
didoEASY+ MR - features added kV measurement
- Dose: 0.2 µGy - 999 Gy (Uncertainty: ±5% @ 25-160 kV / Resolution: 0.01 nGy)
- DLP*: 0.2 µGy*cm - 999 Gy*cm (Uncertainty: ±5% @ 50-160 kV / Resolution: 0.01 nGy*cm)
- Dose Rate: 0.3 µGy/s - 1 Gy/s (Uncertainty: ±5% over full range / Resolution: 0.1 nGy/s)
- Time: 0.5 ms - 300 s (Uncertainty: ± 0,5% / Resolution: 0.1 ms)
- Pulses: 1 - 9999 (Uncertainty: ± 1 pulse)
- Pulse Rate: 0.1 - 250 pulses/s (Uncertainty: ± 0.05 pulses/s)
- kVp / PPV:
- 25 – 35 kV (applies to M detectors)
- 50 – 130 kV (applies to R detectors)
- (Uncertainty: ±5% at calibrated reference points / Resolution: 0.1 kV); Exposure conditions: Mammo: Mo/Mo; HVL 0.6 – 2.2 mm Al eq., correction factors provided for further radiation qualities / R&F: HVL 1.4 – 7.0 mm Al eq.
didoEASY++ MR - features added kV and direct-HVL measurement
- Dose: 0.2 µGy - 999 Gy (Uncertainty: ±5% @ 25-160 kV / Resolution: 0.01 nGy)
- DLP*: 0.2 µGy*cm - 999 Gy*cm (Uncertainty: ±5% @ 50-160 kV / Resolution: 0.01 nGy*cm)
- Dose Rate: 0.3 µGy/s - 1 Gy/s (Uncertainty: ±5% over full range / Resolution: 0.1 nGy/s)
- Time: 0.5 ms - 300 s (Uncertainty: ± 0,5% / Resolution: 0.1 ms)
- Pulses: 1 - 9999 (Uncertainty: ± 1 pulse)
- Pulse Rate: 0.1 - 250 pulses/s (Uncertainty: ± 0.05 pulses/s)
- kVp / PPV:
- 25 – 35 kV (applies to M detectors)
- 50 – 130 kV (applies to R detectors)
- (Uncertainty: ±5% at calibrated reference points / Resolution: 0.1 kV); Exposure conditions: Mammo: Mo/Mo; HVL 0.6 – 2.2 mm Al eq., correction factors provided for further radiation qualities / R&F: HVL 1.4 – 7.0 mm Al eq.
- Direct-HVL: 0.1 – 11 mm Al (HVL output, Uncertainty: ±8% / Resolution: 0.1 mm Al); Internal filtration 0.1 – 12.5 mm Al equivalent (25-35 kV range applies to M detectors / 40-130 kV range applies to R detectors)
* S A Mitchell and C J Martin, Comparison of ionisation chamber and semiconductor detector devices for measurement of the dose–width product for panoramic dental units, J. Radiol. Prot. 33 321 (2013).
Um diese Web App auf Ihrem iPhone/iPad zu installieren, drücken Sie und dann Hinzufügen zum Startbildschirm.